An HPLC assay for two antivirals, rechecked: 315 plates would have done
A 2025 PLOS ONE paper validates a 21-minute RP-HPLC assay for remdesivir and favipiravir. We recompute its retention factors, resolution and plate target with fibub's chromatography maths, and transfer the method to a 3.6-minute run.
Two antiviral drugs, remdesivir and favipiravir, are sometimes packed into the same capsule, and a quality control laboratory needs one assay that measures both. A 2025 paper in PLOS ONE develops and validates such an assay: a plain isocratic reversed-phase HPLC method on a 250 mm column, with favipiravir leaving at 11.5 minutes and remdesivir at 20.95 minutes. The numbers in its system suitability table are exactly the kind that fibub’s chromatography calculator exists to check, and they turn out to describe a separation with enormous room to spare. For the steps on either side, turn peak areas into concentrations with the calibration curve tool and judge the validation figures with method validation statistics.
The paper in brief
The authors ran a C18 column, 250 x 4.6 mm with 5 µm particles, at 1.0 mL/min with a mixture of acetonitrile, methanol and water (50:350:100 by volume) acidified with phosphoric acid, and watched the detector at 300 nm. Injecting 10 µL of a solution containing 0.08 mg/mL favipiravir and 0.04 mg/mL remdesivir gave two peaks, at 11.5 and 20.95 minutes.
Their system suitability limits are the usual pharmacopoeial ones: more than 4,000 theoretical plates for each peak, resolution above 2.0, a tailing factor no greater than 1.5 and a repeatability of 2 % or better (they report 0.38 % and 0.33 %). The validation is thorough: linear from 0.02 to 0.06 mg/mL for favipiravir and 0.022 to 0.066 mg/mL for remdesivir, mean recoveries of 99.76 % and 99.92 %, and forced degradation that destroys 23.6 % of the drug at 70 °C and 17.9 % in acid. The paper lists one weakness of its own method: the run takes a long time.
Checked with the tools
Everything below is a recomputation from the numbers in the paper’s text. Nothing is read off its figures.
Step 1: the dead time and the retention factors. The paper does not report the dead time t₀, so we estimate it from the column: an interstitial plus pore porosity of 0.68 (our assumption, the usual value for a packed C18 column) over a 250 x 4.6 mm tube gives 2.83 mL of mobile phase in the column, which at 1.0 mL/min is t₀ = 2.83 min. From there the tool’s retention factors are k₁ = 3.07 for favipiravir and k₂ = 6.42 for remdesivir, so the selectivity α = 2.09. That is a very large selectivity: the column holds remdesivir more than twice as long as favipiravir.
Step 2: how much resolution 4,000 plates buys. The paper reports resolution only as “more than 2.0”, so we work out what its own efficiency floor implies. Turning N = 4,000 back into a half-height width, W½ = tR√(5.545/N), gives 0.43 min for the favipiravir peak and 0.78 min for the remdesivir peak. The tool’s European Pharmacopoeia resolution from those widths is Rs = 9.2; the more conservative Purnell equation, which assumes both peaks are as wide as the later one, gives Rs = 7.1. Either way the pair is separated four to six times more than the method asks for.
Step 3: the plates this pair actually needs. Run the Purnell equation backwards for a target resolution of 2.0 at this selectivity and retention factor, and the answer is 315 plates. A 250 mm column with 5 µm particles will deliver that in its first two centimetres. The method’s 4,000-plate limit corresponds to a plate height of 62.5 µm, or 12.5 particle diameters, where a healthy modern column runs at 2 to 3: in other words, even a badly damaged column would pass this test, and pass it for good reason, because the separation does not depend on efficiency at all.
| Check | Paper says | The tool gives |
|---|---|---|
| Retention factor, favipiravir | not reported | 3.07 (with t₀ = 2.83 min, our estimate) |
| Retention factor, remdesivir | not reported | 6.42 |
| Selectivity α | not reported | 2.09 |
| Resolution | more than 2.0 | 9.2 from the implied widths, 7.1 from Purnell |
| Plates for Rs = 2.0 | 4,000 required | 315 needed |
| Plate height at 4,000 plates | not reported | 62.5 µm (12.5 particle diameters) |
Step 4: the long run time, fixed. Because the separation is so easy, the method’s only real cost is time and solvent. The tool’s method transfer does the geometric scaling of USP <621> from the paper’s 250 x 4.6 mm, 5 µm column to a modern 100 x 2.1 mm, 1.8 µm one:
- flow 0.579 mL/min instead of 1.0, from the diameter squared and the particle ratio;
- injection 0.83 µL instead of 10 µL, in proportion to column volume;
- every time multiplied by 0.144, so a 25-minute run (our assumption for the run length, since the paper gives only the 20.95 min retention time) becomes 3.6 minutes;
- solvent per injection 2.08 mL instead of 25 mL, that is 8.3 %;
- the length-to-particle-size ratio goes from 50,000 to 55,600, a change of +11.1 %, inside the −25 % to +50 % window USP <621> allows without revalidation of the separation.
The resolution does not suffer: the new column has more plates than the old one, and the pair needed 315.
How we made the graphs
Both charts and every number in this article come from scripts/blog-figures/hplc-remdesivir-favipiravir-plate-count-resolution-method-transfer.mjs in fibub’s public code. It imports retentionFactor, plates, resolution, purnell, platesFor and transfer from src/lib/sci/chrom.ts, the maths behind the chromatography calculator, and prints the values the article quotes.
Inputs from the paper: the column dimensions, the flow, the two retention times, the 4,000-plate limit and the 10 µL injection. Our assumptions, clearly separated: the column porosity of 0.68 that sets the dead time, and a 25-minute run length. The new column size (100 x 2.1 mm, 1.8 µm) is our choice of target, not the paper’s.
How we test the maths
tests/sci.mjs runs on every change to fibub and pins the chromatography module against textbook and pharmacopoeial values:
- a peak at 6.40 min with t₀ = 1.20 min must give k = 4.3333, and peaks at 6.40 and 7.10 min with half-height widths 0.150 and 0.160 min must give Rs = 2.658 (Ph. Eur.) and N = 10,919 for the later peak;
- a pair 1 min apart with 0.5 min baseline widths must give exactly Rs = 2 (USP), and 6,400 plates;
- the Purnell equation and the plates-needed formula must be exact inverses, with 6,272.64 plates needed for Rs = 1.5 at α = 1.1 and k₂ = 5;
- transferring 250 x 4.6 mm, 5 µm at 1.0 mL/min to 100 x 2.1 mm, 1.8 µm must give 0.579 mL/min and a time factor of 0.1444, and a transfer to a 50 mm column must be flagged as outside the USP L/dp window.
New with this article, in tests/sci-tools/chromatography-calculator.mjs: the paper’s own pair must give k₁ = 3.0699, k₂ = 6.4155, α = 2.0898, an implied half-height width of 0.4282 min for the first peak, Rs = 9.21 from those widths, and 314.5 plates needed for a resolution of 2.0.
Check it yourself
On the chromatography calculator page:
- In the first card, type Dead time t₀ (min) 2.83, Peak 1 retention (min) 11.5, Peak 2 retention (min) 20.95.
- Leave Widths measured on at half height (W½) and type Peak 1 width (min) 0.428 and Peak 2 width (min) 0.78.
- Type Column length (mm) 250 and Resolution you need 2.
- You should read retention factors of about 3.07 and 6.42, selectivity 2.09, resolution about 9.2, about 4,000 plates for each peak, a plate height of about 62 µm, and roughly 315 plates needed for your target.
- In the transfer card, type Length now (mm) 250, Inner diameter now (mm) 4.6, Particle size now (µm) 5, Flow now (mL/min) 1, Injection now (µL) 10, then New length (mm) 100, New inner diameter (mm) 2.1, New particle size (µm) 1.8.
- It gives a new flow of 0.579 mL/min, an injection of 0.83 µL, times multiplied by 0.144 and an L/dp change of +11 %, marked as inside the USP window.
What researchers can use it for
- Sanity-check a system suitability table before you submit it: if the plates needed for your target resolution are a few hundred and your limit is thousands, the limit is not protecting the separation and something else should be.
- Shorten a validated method honestly. The transfer card gives the flow, injection and gradient times that keep the chromatogram the same shape, plus the USP L/dp check that says whether the change counts as an adjustment.
- Estimate what you were not given. Retention factors need a dead time; the column volume gives a usable one, and the tool shows how much your conclusions move if the porosity is 0.6 or 0.7 instead.
- Decide where to spend effort. Figure 2 shows that for a pair with α = 1.05 you need more than 100,000 plates for the same resolution; changing the mobile phase or the column chemistry is then the only sensible move.
- Teach it. The same three fields show why a long column is rarely the answer: resolution grows only with the square root of the plates.
The fine print
- The dead time, and so every retention factor and selectivity here, rests on our porosity assumption. The paper reports no t₀ or peak widths.
- Resolution and plate height come from the method’s minimum plate count, not from measured widths, so they are the floor of what the method guarantees, not what the instrument printed.
- The Purnell equation assumes both peaks have the same plate count and uses the width of the later one, which is why it gives a lower resolution than the width-based formula.
- A transfer that changes the column chemistry, not just its size, is a change to the method and needs revalidation, whatever the geometry says.
- The forced degradation and recovery figures are quoted as the paper reports them; we did not recompute them.
Sources
- Khan NU, Iqbal M, Khan BA (2025), A systemic approach to estimate and validate RP-HPLC assay method for remdesivir and favipiravir in capsule dosage form, PLOS ONE 20(4): e0321474, doi:10.1371/journal.pone.0321474. Licence CC BY 4.0, checked on the PLOS article page, 2026-11-24. Values used here: Kromasil 100A C18 column 250 x 4.6 mm, 5 µm; acetonitrile:methanol:water 50:350:100 with 0.5 mL phosphoric acid; 1.0 mL/min isocratic; detection 300 nm; injection 10 µL; retention times 11.5 min (favipiravir) and 20.95 min (remdesivir); more than 4,000 theoretical plates for both peaks; resolution more than 2.0; tailing not more than 1.5; linearity 0.02 to 0.06 mg/mL (favipiravir) and 0.022 to 0.066 mg/mL (remdesivir); mean recovery 99.76 % and 99.92 %.
- Plate count, resolution, tailing and the geometric method-transfer rules: USP general chapter <621> Chromatography and the European Pharmacopoeia general chapter 2.2.46.
- Purnell, J. H. (1960), Nature 184: 2009 (the resolution equation in terms of plates, selectivity and retention factor).
If you are an author or publisher and want a credit or a passage changed or removed, tell us and we will act quickly.