Home / Science / Physics / Nuclear and particle physics / Counting statistics helperTool #950

Radiation counting statistics: net count rate, Currie detection limit, MDA, dead time and counting time

For Geiger, scintillation and gamma counting: subtract the background with the right uncertainty, decide whether anything was detected (Currie's decision threshold L_C and detection limit L_D), get the activity and minimum detectable activity, exact Poisson limits, dead-time corrections for both detector models, and the best split of counting time.

Keep it one tap away for everyday use.

Bookmark Counting statistics helper?

Press Ctrl + D now. Your browser opens its own bookmark box, where you choose the name and folder.

  1. Counts
  2. Net rate
  3. Detected?
  4. Dead time
Nothing is uploadedWorks offline once open

1. Net counts and detection limit

Efficiency: counts recorded per particle emitted (from a calibration source); yield: particles of that kind per decay. Leave it blank to work in counts only. The example numbers are made up.

2. Dead-time correction

Non-paralysable: the detector ignores events while it is busy. Paralysable: every event, even one it misses, restarts the busy time (so very high rates can read low). Real detectors sit between the two.

3. Plan the counting time

Poisson counting statistics only: add calibration, geometry and other uncertainties yourself. For study and planning, not a substitute for professional judgement or your laboratory's validated procedures.

Download

What it does

Radioactive decay is random, so every count has an uncertainty: for N counts the standard deviation is √N. This page does the everyday sums of a counting lab with that in mind.

  1. Net counts and detection limit. From gross counts (sample plus background) and a separate background count, it gives the net rate with its uncertainty, √(G/t_g² + B/t_b²), and Currie’s two limits. The decision threshold L_C is the net count above which you say “detected”, with a false-positive rate α. The detection limit L_D is the true net count you would detect with probability 1 − β. With a counting efficiency it also gives the activity and the minimum detectable activity (MDA). Exact Poisson limits are shown for the gross and background counts.
  2. Dead-time correction. A detector is blind for a short time τ after each event. The page corrects a measured rate with the non-paralysable model, n = m / (1 − mτ), and the paralysable one, m = n e^(−nτ) (solved with the Lambert W function), and plots both.
  3. Counting-time planning. For expected gross and background rates, the best split of a fixed time is t_sample / t_background = √(R_gross / R_background); the page gives that split, the uncertainty it reaches, and the total time for the uncertainty you want.

How to use it

Type the counts and times (in seconds, minutes or hours; the same unit for both), and optionally the efficiency (counts per particle emitted) and the emission yield (particles per decay). Pick α and β (5 % each is usual) and the confidence for the intervals. The other two cards work on their own.

Worked example

A made-up sample gives 1,500 counts in 600 s; the background gives 900 counts in 1,800 s. The net rate is 2 ± 0.067 per s, 3.33 %. With no activity, the net count in 600 s would scatter with σ₀ = 20 counts, so L_C = 32.9 and L_D = 68.5 net counts; the 1,200 net counts are far above L_C, so the page says detected. At 25 % efficiency the activity is 8 ± 0.27 Bq and the MDA is 0.4567 Bq (27.4 dpm).

A counter reading 20,000 per s with a 5 µs dead time is busy 10 % of the time: the true rate is 22,222 per s (non-paralysable) or 22,367 per s (paralysable). With rates of 2.5 and 0.5 per s and 2,400 s in all, count the sample for 1,658 s and the background for 741.6 s to reach 2.34 %; reaching 1 % needs 13,090 s.

Tips

  • Count the background for at least as long as the sample when the sample is weak; the split rule tells you how much longer.
  • L_C is for deciding about this measurement; L_D (and the MDA) describe what the method can detect, before you measure.
  • Keep dead-time losses below about 10 % when you can; the two models start to disagree above that.

FAQ

Why do two “detected” numbers exist? L_C controls false alarms; L_D controls missed detections. A true signal exactly at L_C would be missed half the time, so L_D is about twice as large.

Is the normal approximation good enough? For a few tens of counts or more, yes. For very small counts the exact Poisson limits shown for the gross and background counts are safer.

Is this enough for a regulatory report? No. It covers Poisson statistics only. It is for study and planning, not a substitute for professional judgement or your laboratory’s validated procedures.

Where the numbers come from

  • Currie, L. A. (1968), Limits for qualitative detection and quantitative determination, Analytical Chemistry 40, 586 to 593 (decision threshold and detection limit formulas).
  • Poisson counting, dead-time models and optimum counting-time division: standard results, for example Knoll, G. F., Radiation Detection and Measurement. Exact Poisson limits: Garwood, F. (1936), Biometrika 28, 437.
  • Checked against SciPy (normal and chi-square quantiles, Lambert W, root finding) and against Currie's paired-blank forms L_C = 2.33 √B and L_D = 2.71 + 4.65 √B, in tests written for this page.
  • The example counts are made up.

Standard scientific values, such as constants, atomic weights and melting points, are facts found in every textbook; we name where ours were checked. If you are an author or publisher and want a value, a credit or a table changed or removed, tell us and we will act quickly.

Last updated 2026-11-09.

Related tools

Decay chain simulator #945 Follow a radioactive parent and up to seven daughters through time with the Bateman equations: activity of every member at any time, when each daughter peaks, whether the chain reaches secular or transient equilibrium and how long that takes. Branching, any starting activities, and a chart and CSV of the ingrowth. Radioactive decay and specific activity calculator #99 Activity of F-18, Tc-99m, I-131, C-14, P-32 and 28 more nuclides (or any half-life) at any date and time, a printable decay table, when it falls to a given activity, carrier-free specific and molar activity, and the labelled fraction of your sample. Bq, MBq, GBq, µCi, mCi and Ci. Gamma shielding calculator #946 How much a lead, tungsten, steel, aluminium, concrete or water shield cuts a gamma beam from 0.1 to 2 MeV: linear attenuation coefficient, half- and tenth-value layers, narrow-beam transmission, optional buildup, and the thickness for a wanted transmission, with every material compared. The built-in μ/ρ is worked out from formulas, not a table, so it is only an estimate: type the NIST XCOM value for real work. Spectrum peak fitter #951 Paste or open a spectrum or histogram (counts per channel), mark a region, find or type the peaks, and fit Gaussians on a constant, straight or quadratic background with Poisson weights: centroid, FWHM, net area and their uncertainties, χ²/ν and residuals. Then calibrate channel to energy from known peaks. Uncertainty propagation calculator #25 Type your formula and your measurements with their uncertainties, or paste repeated readings. Get the result with its uncertainty, which input matters most, and a Monte Carlo check. Relativistic kinematics calculator #947 Four special-relativity calculators in one page, in MeV or GeV with c = 1: the invariant mass of any set of particles, the momenta and lab energy ranges in a two-body decay, a Lorentz boost of a four-momentum in any direction, and the threshold energy of a reaction on a fixed target or in a collider. Type particle names instead of masses.