Electron diffraction indexer: camera constant, ring ratios, d-spacings and zone axis for TEM patterns

Index TEM and SAED patterns in the browser: the relativistic electron wavelength and camera constant λL, calibration with rings of a known standard, d-spacings from ring radii or spot distances, ring-ratio indexing of cubic lattices with the lattice parameter, and the zone axis of a two-spot pattern.

Keep it one tap away for everyday use.

Bookmark Electron diffraction indexer?

Press Ctrl + D now. Your browser opens its own bookmark box, where you choose the name and folder.

  1. Microscope
  2. Calibrate
  3. Rings
  4. hkl and a
Nothing is uploadedWorks offline once open

Microscope

Rd = λL holds for the small angles of electron diffraction (2θ is a degree or two). The nominal camera length on the microscope is often a few per cent off, so calibrating with a standard of known lattice parameter, under the same lens settings, is more accurate. With distances in 1/nm, d = 1/R and no camera constant is needed.

Calibrate with a standard

Use the lattice parameter from your standard's certificate. The example standard is a made-up face-centred cubic film with a = 4.08 Å, about what a gold film gives.

Rings of your sample

What it does

In a transmission electron microscope, a polycrystalline sample gives rings and a single crystal gives spots. Every ring or spot at distance R from the centre comes from planes of spacing d, through the camera equation R d = λ L: λ is the electron wavelength and L the camera length, and their product is the camera constant. This page does the whole chain:

  • Wavelength. λ from the accelerating voltage, with the relativistic correction, which matters: at 200 kV the electrons travel at 0.70 c and λ is 2.508 pm rather than 2.742 pm.
  • Camera constant. λL from the voltage and the nominal camera length, or better, from a calibration: rings of a standard with a known lattice parameter are indexed and λL = R d(hkl) is averaged over them. The page shows how far the nominal value was off.
  • d-spacings. From radii or diameters in mm, pixels (with a pixel size or a calibration in pixels) or 1/nm from a calibrated image. An exact-geometry option uses tan 2θ = R/L instead of the small-angle rule.
  • Ring-ratio indexing. For a cubic lattice R² is proportional to N = h² + k² + l². The page tries simple, body-centred, face-centred and diamond cubic, gives hkl for each ring and the lattice parameter a, and lists allowed rings that are missing.
  • Spot patterns. Give two spot distances and the angle between them. The page finds the hkl pairs of the lattice that fit and their zone axis, the direction the beam runs along.

How to use it

  1. Microscope. Type the voltage and camera length, and pick the units you measured in and whether you measured radii or diameters.
  2. Calibrate. Enter the standard’s lattice type and a, and its ring radii, measured under the same conditions. Set Camera constant from to the standard. Skip this to use the nominal value.
  3. Sample. Enter your ring radii, smallest first. Read the lattice, a and the hkl of each ring.
  4. Spots. For a single crystal, enter two spot distances and their angle to get the zone axis.
  5. Download the ring table as CSV, or a PNG, PDF or Word report.

Worked example

Press Use the example. At 200 kV, λ = 2.5079 pm and the nominal camera constant for L = 1000 mm is 25.079 Å·mm. The made-up standard’s seven rings (10.69 to 26.87 mm) index as 111, 200, 220, 311, 222, 400 and 331, giving a calibrated λL = 25.092 ± 0.081 Å·mm, 0.05 % above the nominal.

The sample’s eight rings, 12.31 to 34.99 mm, have R² ratios 1 : 2.03 : 3.01 : … : 8.08, which only body-centred cubic explains without missing rings: body-centred cubic, a = 2.872 Å, with rings 110, 200, 211, 220, 310, 222, 321 and 400. Face-centred cubic could only fit them by leaving out rings it should show.

The two example spots at 12.36 and 17.48 mm, 45° apart, are 110 and 020 of that lattice, and the zone axis is [001].

Tips

  • Measure diameters across the pattern and halve them: it removes the error in finding the centre.
  • Calibrate at the same voltage, camera length, lens currents and specimen height as the sample.
  • Ring-ratio indexing needs the first rings. If a weak first ring is missing, a lattice may fit with a different first index; check the “not seen” column.
  • Simple cubic and body-centred cubic give identical ratios up to the sixth ring; the seventh decides.

FAQ

Why is the relativistic correction needed? At TEM voltages electrons move at a large fraction of the speed of light. Without the correction λ is about 9 % too long at 200 kV and every d would be wrong by as much.

Can I index non-cubic patterns? Not automatically. Use the d-spacings from the table with the Miller index calculator, which gives d for any cell.

Are my measurements uploaded? No. Everything is worked out on this page.

Where the numbers come from

  • The relativistic electron wavelength uses CODATA 2018 constants; the camera equation Rd = λL, ring-ratio indexing and the zone law are standard textbook electron microscopy. The code is written by fibub and checked in the tests against scipy.constants and NumPy.
  • The example rings are made up by the page: a face-centred cubic standard with a = 4.08 Å and a body-centred cubic sample with a = 2.87 Å at 200 kV and 1000 mm, with small made-up measuring errors.

Standard scientific values, such as constants, atomic weights and melting points, are facts found in every textbook; we name where ours were checked. If you are an author or publisher and want a value, a credit or a table changed or removed, tell us and we will act quickly.

Last updated 2026-11-09.

Related tools

Molecular point group finder #110 Paste a molecule's XYZ coordinates or open an .xyz file to get its point group (C₂ᵥ, D₃h, T_d, O_h, I_h and the rest), every symmetry element, the group order, and whether the molecule can be chiral or polar. No coordinates? Answer the flowchart questions instead. 23 example molecules. XRD peak calculator #842 Turn powder X-ray diffraction peaks into d-spacings, crystallite sizes (Scherrer, with the instrument's width taken away), size and microstrain from a Williamson-Hall plot, and a cubic lattice (simple, body-centred, face-centred or diamond) with its lattice parameter. Type peaks or open a 2θ scan and let the page find and fit them. CIF viewer #843 Open a Crystallographic Information File (CIF) and see the unit cell in 3D: the symmetry operations fill the cell from the listed atoms, and the page gives bond lengths and coordination, the cell volume, the formula and formula units Z, and the density. The file never leaves your browser. Miller index calculator #844 Work out the spacing of any (hkl) plane in cubic, tetragonal, orthorhombic, hexagonal, trigonal, monoclinic or triclinic cells, the angle between two planes, their zone axis and the 2θ of each, convert to and from four-index Miller-Bravais notation, list reflections, and see the planes drawn in the unit cell. 1D Schrödinger solver #923 Choose a textbook potential, type a formula or draw your own, and get the lowest energy levels and wavefunctions of the one-dimensional Schrödinger equation by the matrix (finite-difference) method, with the exact levels alongside for the box, harmonic oscillator, finite well, Morse, linear and Pöschl–Teller potentials. Adduct and neutral loss calculator #984 Type a formula or a neutral monoisotopic mass and get the m/z of 25 common electrospray adducts in positive and negative mode, the ions left after common neutral losses (water, ammonia, CO2, sugars, phosphate and your own), and match a list of observed m/z values within a ppm tolerance. With no molecule, it turns observed m/z values into candidate neutral masses.